MIXED FLOWING GAS

 

ABOUT

 
 
Take a tour of Mixed Flowing Gas: 1 | 2 | 3 | 4 | 5 | 6

Section of MFG lab

Mixed Flowing Gas testing requires specific instrumentation to meet most testing standards. A primary requirement is a chamber that maintains a constant 30º C at a relative humidity of 70 to 75%, the standard for most  tests.  The chamber needs to maintain a temperature and humidity tolerance of about +/-1º C and +0, -2% relative humidity in order to avoid inconsistent results that can occur past 70% relative humidity.  A rise in sulfur corrosion at relative humidity levels above 70% is typically the problem.